故障现象:扫描中断故障现象分析:经仔细观察,扫描中断大多发生在早上九点钟左右,都是在连续做了几个增强后,扫描中断。但经过大约一个小时后,系统又可以扫描。经过测量及监控,排除了供电电源问题。继续软件测试,排除系统软件故障,根据故障现象判断不是球管打火损坏(球管打火损坏后,系统不会再重新可以扫描)。进入系统报告,在扫描中断发生的时候,球管油温都在79℃~80℃。浏览历史记录,正常情况下系统温度很少超过55℃,因此怀疑为球管冷却问
Symptoms: Scan interrupt failure Phenomenon analysis: After careful observation, scan interruption occurred mostly in the morning at about nine o’clock, are doing a few enhancements in a row, the scan interrupt. But after about an hour, the system can scan again. After measuring and monitoring, ruled out the power supply problems. Continue software testing to rule out the system software failure, according to the failure to determine the ball is not damaged tube ignition (tube damage after the fire damage, the system will not be able to re-scan). Into the system report, the interruption occurred in the scan, the tube oil temperature at 79 ℃ ~ 80 ℃. Browsing history, under normal circumstances, the system temperature rarely exceeds 55 ℃, so it is suspected that the cooling of the tube asked