论文部分内容阅读
A small-angle scattering neutron spectrometer for material research is under construction at the China Spallation Neutron Source. An intervening neutron beam monitor behind the sample is needed to measure the beam intensity in order to reduce the measurement uncertainty caused by beam fluctuation. Considering the mobility requirement and limited space, we proposed a compact monitor using a type of lithium-glass scintillator provided by China Building Materials Academy. Its performance was studied experimentally using 252Cf and 60Co sources. The neutron light yield of the selected scintillator was measured to be 5.3 × 103 photons/neutron. The feasibility of n-gamma discrimination using the charge comparison method was verified. By using the Geant4 toolkit, the monitor was modeled with precise physical processes including neutron tracking, scintillation, and optical photon transmission. The gamma sensitivity and detection effi-ciency were investigated in the simulation. It was con-cluded that a 0.5-mm-thick lithium-glass scintillator with natural lithium is an appropriate choice to satisfy both the neutron detection efficiency and gamma elimination requirements.