本文研究了声图像的恢复对声图像纹理分析参数的影响。论文阐述了一种用于B超图像恢复的方法、它利用三个声学特性已知的超声组织仿真模块作参考;在增益与TGC设置不变的情况下,可对均匀组织的B超图像进行恢复。论文采用了几种不同的纹理分析方法对已恢复的B超图像及未经恢复旦在不同设置下采集的图像进行了分析结果的对比研究,得到了一些有实用意义的结论。
This paper studies the effect of acoustic image restoration on acoustic image texture analysis parameters. In this paper, a method for B-mode image restoration is described, which uses three known ultrasonic tissue simulation modules for acoustic reference. In the case of constant gain and TGC settings, B-mode images can be uniformly organized restore. The paper uses several different texture analysis methods to compare the recovered B-images and the images collected under different settings without restoration. Some useful conclusions have been obtained.