A rapid method to determine glutamic acid (Glu) in cerebrospinal fluid (CSF) by capillaryelectrophoresis with high frequency conductivity detection (contactless conductivity detection) wasdescribed. The CSF sample was pretreated with silver cation resin to remove high concentration ofCl- ions in CSF. The separation was achieved in the buffer solution of 10 mmol/L Tris and 8mmol/L boric acid at the separation voltage of 20.0 kV. Glu showed linear response in the range of5.0×10-6 to 6.0×10-3 mol/L, the limit of detection was 1.0×10-6 mol/L. The method was used foranalysis Glu in CSF satisfactorily with a recovery of 97.8-98.8%.
A rapid method to determine glutamic acid (Glu) in cerebrospinal fluid (CSF) by capillary electrophoresis with high frequency conductivity detection (contactless conductivity detection) wasdescribed. The CSF sample was pretreated with silver cation resin to remove high concentration of Cl- ions in CSF. The separation was achieved in the buffer solution of 10 mmol / L Tris and 8 mmol / L boric acid at the separation voltage of 20.0 kV. Glu showed linear response in the range of 5.0 × 10 -6 to 6.0 × 10 -3 mol / L , the limit of detection was 1.0 × 10 -6 mol / L The method was used foranalysis Glu in CSF satisfactorily with a recovery of 97.8-98.8%.