论文部分内容阅读
CT是一种先进、精密的大型医疗设备,它不仅为医学影像诊断提供了重要的依据,而且也是保证医院经济效益的重要基础。因此正确的使用,精心的维护保养,以确保机器开机率是及其重要的。本文介绍了关于Brightspeed16排螺旋CT在使用当中的故障现象,以及诊断分析和解决方法。1故障现象GE公司Brightspeed型号的16排CT,球管预热过程正常,但是做每日校准或做病人时,报错:“error occurred attempting to home CAM B or CAM B failed to move to the correct position for beam tracking”。2诊断过程
CT is an advanced and sophisticated large-scale medical device, which not only provides an important basis for medical image diagnosis, but also serves as an important foundation for ensuring the economic benefits of the hospital. Therefore, the correct use, careful maintenance, to ensure that the machine is a breeze rate and its important. This article describes the symptoms of using the Brightspeed 16-Row Spiral CT in the field, as well as diagnostic analysis and solutions. 1 Symptom GE’s Brightspeed model of 16-row CT, the preheating tube process is normal, but doing daily calibration or patient, an error: “The occurrence of CAM B or CAM B failed to move to the correct position for beam tracking ”. 2 diagnostic process