作者最近设计出二种新型非心外膜电极系统。用一支12F 右心三极导管,从锁骨下静脉插入,其顶端电极发出起搏信号并可记录;另有二个相距12cm 的复律/除颤电极,远端电极长4cm,面积为400mm~2,近端电极长8cm,面积为800mm~2。三个电极分别置于右室尖、右室和上腔静脉右房连接处。方法Ⅰ以远端电极为阴极,以近端电极为阳极。方法Ⅱ以远端电极为阴极,另以一个表面积为50cm~2的电极贴在胸壁心尖部皮肤为阳性。研究对象为12例冠心病患者,心功能均为 NYHA Ⅲ、Ⅳ级,临床均有不明原因的
The authors recently designed two new non-epicardial electrode systems. With a 12F right heart catheterization, inserted from the subclavian vein, the apical electrode sends a pacing signal and can be recorded; there are two cardioversion / defibrillation electrodes 12 cm apart and the distal electrode is 4 cm long and has an area of 400 mm ~ 2, the proximal electrode length 8cm, an area of 800mm ~ 2. Three electrodes were placed in the right ventricular apex, right ventricular and superior vena cava right atrium junction. Method I to the distal electrode as the cathode, the proximal electrode as the anode. Method Ⅱ to the distal electrode as the cathode, and the other with a surface area of 50cm ~ 2 electrodes attached to the chest wall apical skin was positive. The study object was 12 patients with coronary heart disease, cardiac function were NYHA Ⅲ, Ⅳ level, there are clinically unknown reasons