论文部分内容阅读
本文运用程序胸壁刺激试验对33例置入按需起搏器的感知功能进行检查,发现不同型号的起搏器感知外电刺激的全抑制区是不同的;测出的起搏器不应期也与其额定值有所差别,但变化仍在正常范围内。该试验方法简便易行,适于一般医院对置入按需起搏器患者进行安全性监测。
In this study, chest wall stimulation test procedures were used to detect 33 patients with on-demand pacemaker sensing and found that different types of pacemaker perception of external electrical stimulation of the total inhibition zone is different; measured pacemaker refractory period also And its rating vary, but the change is still within the normal range. The test method is simple and convenient, suitable for general hospital on-demand pacemaker patients for safety monitoring.