论文部分内容阅读
In this study, a compact 16-channel integrated charge-sensitive preamplifier named the smart preamplifier (SPA) was developed to support the large-scale detector array used in mode nuclear physics experiments. Two types of SPA, namely SPA02 and SPA03 (with extealfield effect transistor), have been manufactured to match silicon detectors with small and large capacitances, respectively. The characteristics of the SPA include fast response of typically less than 6 ns for pulse rising time and low equivalent noise of 1.5 keV at zero input capacitance. The energy sensitivity and pulse decay time can be easily adjusted by changing the feedback capacitance Cf and resistance Rf in various applications. A good energy reso-lution of 24.4 keV for 5.803-MeV alpha particles from 244Cm was achieved using a small-sized Si-PIN detector;for the silicon strip detectors in the test with the alpha source, a typical energy resolution of 0.6–0.8% was achieved. The integrated SPA has been employed in sev-eral experiments of silicon strip detectors with hundreds of channels, and a good performance has been realized.