体检车可作为小型流动体检中心,为企事业单位上门提供体检服务。但车载X光机的故障率较高,影响体检工作的顺利进行。我们通过安装电源保护器,使车载X光机及其他设备的故障率明显降低,同时降低了设备维修成本。1车载X光机常见故障1.1电气故障电路的保险、继电器、电容容易损坏。因为多数企业没有为X光机准备专用线路,X光机多与空调、车内设备共用一条线路;另外,外接电源接线固定不好、接触
Medical examination car can be used as a small mobile medical examination center, for medical institutions to provide medical services. However, the high failure rate of X-ray car, affecting the smooth progress of the medical examination. By installing a power protector, we have significantly reduced the failure rate of X-ray machines and other equipment in the vehicle while reducing equipment maintenance costs. 1 car X-ray machine common fault 1.1 electrical fault circuit insurance, relays, capacitors easily damaged. Because most companies do not have a dedicated line for the X-ray machine, X-ray machine and air conditioning, car equipment sharing a line; In addition, the external power wiring is not fixed, contact