论文部分内容阅读
Prospeed AI CT故障原因,主要分两类。一类是误操作引起的,当机器不正常时首先要考虑是否有误操作,以免走弯路。第二类是软、硬件故障,硬件损坏时多伴有故障提示,根据情况具体分析,容易找准故障点。1误操作引起的故障(1)故障现象一扫描架不转,机器无法正常工作,重复开关机几次,故障依旧。根据扫描架上的指示灯提示,是机器被紧急停机,发现是床尾急停开关不小心被按下,造成此假故障。恢复急停开关位置,按控制台键盘上的复位键后设备正常。分析:床边及门架的紧急停机开关,容易被病人配属及工
Prospeed AI CT causes of failure, mainly divided into two categories. One is caused by misuse, when the machine is not normal, we must first consider whether there is a mistake, to avoid detours. The second category is software and hardware failures, hardware failure accompanied by more tips, according to the specific analysis, easy to find the right point of failure. 1 Malfunction caused by misuse (1) Failure phenomenon A gantry does not turn, the machine does not work, repeat the switch a few times, the fault still. According to the gantry indicator light prompts, the machine was an emergency stop, found to be emergency stop button bed accidentally pressed, causing this false failure. Restore emergency stop switch position, press the reset button on the console keyboard after the device is normal. Analysis: bedside and mast emergency shutdown switch, easily attached to the patient and work