目的使用激光多普勒血流仪(LDF)比较人的牙髓、手指、前臂、牙龈血流信号,评估血流仪测量牙髓血流的能力。方法使用激光多普勒血流仪对28例受试者的指尖、前臂、牙龈、不透明的橡皮障覆盖下的上颌中切牙以及去除橡皮障后的中切牙血流信号进行测量,记录测量结果 ,并进行统计分析。结果所有测量区域中血流信号在手指最大,依次是牙龈,没有橡皮障的牙齿,前臂,有橡皮障的牙齿。使用不透明橡皮障的牙齿的血流信号接近0。结论实验结果显示人的牙髓血流信号非常低,血流仪对于牙髓血流的检测有一定的限制。
Objective To compare the blood flow of human dental pulp, finger, forearm and gingiva with laser Doppler flowmetry (LDF) to evaluate the ability of the blood flow meter to measure the flow of dental pulp. Methods Laser wave Doppler flowmeter was used to measure the signal of the maxillary central incisors and the incisors of the central incisors after the removal of the rubber cochlea in 28 subjects’ fingertips, forearm, gingiva, opacity Measurement results, and statistical analysis. Results The blood flow signals in all measurement areas were greatest at the fingers, followed by gums, gum-free teeth, forearms, rubber-damped teeth. The blood flow signal of the teeth using the opaque rubber barrier is close to zero. Conclusion The experimental results show that the human dental pulp blood flow signal is very low, and the blood flow meter has some limitations on the detection of pulp blood flow.