目的了解医用诊断X射线摄片设备性能状况,确保医疗安全。方法按照《医用X射线诊断设备影像质量控制检测规范》WS/T189-1999进行检测。结果检测医用诊断X射线摄片设备93台,合格38台,合格率仅40.86%,特别是乡镇医院检测医用诊断X射线摄片设备57台,合格6台合格率仅10.53%。结论应加强医用诊断X射线机状态检测工作。以保证医疗质量和安全,并加大投入,加快乡镇医院医用诊断X射线摄片设备的更新换代工作。
Objective To understand the performance of medical diagnostic X-ray equipment to ensure medical safety. Methods According to “Medical X-ray diagnostic equipment image quality control testing specification” WS / T189-1999 for testing. Results A total of 93 sets of medical diagnostic X-ray equipment were detected, of which 38 were qualified, with a pass rate of only 40.86%. In particular, 57 sets of medical diagnostic X-ray radiography equipment were tested at the township hospitals, passing 6 passes with a pass rate of 10.53%. Conclusion The diagnosis of medical diagnostic X-ray machines should be strengthened. To ensure the quality and safety of medical treatment, and increase investment, speed up the township hospital medical diagnostic X-ray equipment replacement.