气相色谱仪是教学常用设备,本文就该仪器几种常见故障及维修情况介绍如下。1基线异常1.1故障现象1基线走不稳机型:SP一2305故障分析及维修:该现象发生于久未开机。先断开同轴电缆与放大器输入级的连接,走放大器基线,基线稍稍变稳,但仍有浮移。再从主机上卸下?
Gas chromatograph is commonly used in teaching equipment. This article describes several common faults and maintenance of the instrument as follows. 1 Baseline anomaly 1.1 Fault phenomenon 1 Baseline is unstable: SP-2305 fault analysis and maintenance: This phenomenon occurs when the power is turned on for a long time. Disconnect the coaxial cable from the input stage of the amplifier and take the amplifier baseline. The baseline is slightly stable but still floating. Remove from the host?