防护热板法装置热板功率损失的补偿方法

来源 :纳米技术与精密工程 | 被引量 : 0次 | 上传用户:Wangqiling1116
下载到本地 , 更方便阅读
声明 : 本文档内容版权归属内容提供方 , 如果您对本文有版权争议 , 可与客服联系进行内容授权或下架
论文部分内容阅读
在防护热板法装置中,沿整个隔缝的温度分布是不均匀的,导致热板与护板间的功率损失,影响导热系数测量的准确度.针对该问题提出一种补偿方法,通过计算热板功率损失修正热板功率,使修正后热板的功率准确反映理想一维条件下流过试件的热流量,从而提高导热系数测量准确度.实验结果显示,通过提出的补偿方法测得的导热系数准确度达到±2!,有效地解决了由于隔缝温度不平衡而造成的测量导热系数不准确的问题. In the protection of hot plate method, the temperature distribution along the entire gap is not uniform, resulting in the power loss between the hot plate and the protective plate, affecting the accuracy of the thermal conductivity measurement. A compensation method is proposed for this problem, Hot plate power loss correction hot plate power, so that the corrected hot plate power accurately reflects the ideal one-dimensional flow through the specimen under the heat flow, thereby improving the thermal conductivity measurement accuracy.The experimental results show that the compensation method by the proposed Thermal conductivity accuracy of ± 2 !, Effectively solve the gap due to the temperature imbalance caused by the measurement of thermal conductivity inaccurate problem.