目的探讨CT增强扫描检查中预热造影剂与造影剂相关不良反应发生的相关性,为减少影像检查造影剂相关不良反应的发生提供参考。方法选择2015年7月至2016年8月某院1786例行多层螺旋CT检查的患者随机分为实验组(n=910)及对照组(n=876),实验组将造影剂提前预热至37℃后进行造影剂注射剂CT增强检查,对照组则常温处理,比较两组不良反应事件的发生率。结果实验组发生不良反应事件共8例,其中轻度、中度、重度分别为6、2、0,发生率为0.88%,药物外渗0例;而对照组发生不良反应事件为26例,其中轻度、中度、重度分别为21、4、1,发生率为2.97%,药物外渗2例;实验组的发生率明显低于对照组,有统计学差异P<0.05。结论预热造影剂至37℃能明显降低CT增强扫描中的不良反应事件发生,可在临床应用中广泛推广。
Objective To investigate the correlation between preheated contrast media and contrast-related adverse reactions in CT-enhanced scanning and to provide reference for reducing the incidence of adverse effects of contrast-enhanced imaging. Methods From July 2015 to August 2016, 1786 patients underwent multi-slice spiral CT in a hospital were randomly divided into experimental group (n = 910) and control group (n = 876). The experimental group preheated the contrast medium To 37 ℃ contrast agent injection CT enhanced examination, the control group was treated at room temperature, the incidence of adverse reactions in both groups were compared. Results There were 8 cases of adverse reactions in the experimental group, of which mild, moderate and severe were 6, 2 and 0 respectively, the incidence rate was 0.88% and the drug extravasation was 0 cases. In the control group, there were 26 cases of adverse reactions, The mild, moderate and severe cases were 21,4,1 respectively, the incidence was 2.97% and the drug extravasation was 2 cases. The incidence of experimental group was significantly lower than that of the control group (P <0.05). Conclusion Preheating the contrast agent to 37 ℃ can significantly reduce the incidence of adverse reactions in CT enhanced scanning and can be widely used in clinical practice.