A low power low noise analog front end for portable healthcare system

来源 :Journal of Semiconductors | 被引量 : 0次 | 上传用户:dsclq
下载到本地 , 更方便阅读
声明 : 本文档内容版权归属内容提供方 , 如果您对本文有版权争议 , 可与客服联系进行内容授权或下架
论文部分内容阅读
The presented analog front end(AFE) used to process human bio-signals consists of chopping instrument amplifier(IA),chopping spikes filter and programmable gain and bandwidth amplifier.The capacitorcoupling input of AFE can reject the DC electrode offset.The power consumption of current-feedback based IA is reduced by adopting capacitor divider in the input and feedback network.Besides,IA’s input thermal noise is decreased by utilizing complementary CMOS input pairs which can offer higher transconductance.Fabricated in Global Foundry 0.35 m CMOS technology,the chip consumes 3.96 A from 3.3 V supply.The measured input noise is 0.85 Vrms(0.5–100 Hz) and the achieved noise efficient factor is 6.48. The presented analog front end (AFE) used to process human bio-signals consists of chopping instrument amplifier (IA), chopping spikes filter and programmable gain and bandwidth amplifier. The capacitor coupling input of AFE can reject the DC electrode offset. The power consumption of current-feedback based IA is reduced by adoption capacitor divider in the input and feedback network.Besides, IA’s input thermal noise is decreased by utilizing complementary CMOS input pairs which can offer higher transconductance. Fabrication in Global Foundry 0.35 m CMOS technology, the chip consumes 3.96 A from 3.3 V supply. The measured input noise is 0.85 Vrms (0.5-100 Hz) and the achieved noise efficient factor is 6.48.