背景:种植体及种植体周围骨应力过大是影响种植体长期使用的主要因素之一,因此种植体外形的科学设计与选择是延长种植体使用寿命的重要手段之一。目的:对课题组前期开发的可任意调节角度牙种植体中的连接体螺钉和中央螺钉直径进行优化分析与设计。方法:使用Pro/E 5.0、Mimics 10.0、ANSYS Workbench 14.5软件建立含有可调节角度牙种植体的无牙牙合下颌骨有限元模型,分析种植体-骨组织系统的最大等效应力。结果与结论:从灵敏度曲线可以看出,种植体-骨组织系统的最大等效应力受连接体螺钉直径变化的影响颇大,而受中央螺钉直径变化的影响可以忽略不计。当施加垂直向力时,连接体螺钉直径为1.29 mm、中央螺钉直径为0.95 mm时,种植体-骨组织系统最大等效应力最小;当施加水平舌向力时,连接体螺钉直径为1.45 mm、中央螺钉直径为1.09 mm时,种植体-骨组织系统最大等效应力最小。结果表明连接体螺钉直径对下颌骨和种植体-基台复合体等效应力峰值的影响远大于中央螺钉直径,提示在种植体选择和设计中相对于中央螺钉而言应更重视连接体螺钉的选择和设计。
BACKGROUND: Bone stress around implants and implants is one of the main factors that affect the long-term use of implants. Therefore, the scientific design and selection of implants are one of the important means to prolong the service life of implants. OBJECTIVE: To optimize the design and analysis of the diameter of the connector screw and the central screw in the dental implant which can be adjusted at an earlier stage. Methods: The maxillary equivalent stress of the implant - bone tissue system was analyzed by using Pro / E 5.0, Mimics 10.0 and ANSYS Workbench 14.5 software to establish a toothless occlusal mandibular finite element model with adjustable angle dental implants. RESULTS AND CONCLUSION: From the sensitivity curve, it can be seen that the maximum equivalent stress of implant-bone system is greatly influenced by the change of screw diameter of the connector, but the influence of the change of the diameter of the central screw is negligible. When the vertical force was applied, the maximum equivalent stress of the implant-bone tissue system was the smallest when the screw diameter of the connector was 1.29 mm and the diameter of the central screw was 0.95 mm. When the horizontal lingual force was applied, the screw diameter of the connector was 1.45 mm When the diameter of the central screw is 1.09 mm, the maximal equivalent stress of the implant-bone tissue system is the smallest. The results show that the effect of screw diameter on the equivalent stress peak of the mandibular and implant-abutment complex is much greater than that of the central screw, suggesting that more emphasis should be placed on the ratio of central screw to that of the central screw in implant selection and design Choice and design.