培养液体菌种的摇床间,需要较准确的室温,但一般窗式空调机控温精度较差,误差达2~3℃,影响了实验的准确性。要在不改动空调机的情况下使室温稳定在±0.5℃范围,可采取如下方法:找一台电子继电器(如7501型),一支0~50℃电接点水银温度计(如WXG型),一个和空调机相配的插座,安装在一块电插板上,并接好线。因为空调机停机2分钟后方可再开机,所以为使电接点水银温度计不过于太灵敏,应在其感温部分套上一根2厘米长的输血用的乳胶管。
The cultivation of liquid bacteria shaking room, need more accurate room temperature, but the general window air conditioner temperature control accuracy is poor, the error of 2 ~ 3 ℃, affecting the accuracy of the experiment. To stabilize the room temperature within ± 0.5 ° C without changing the condition of the air conditioner, take the following measures: Find an electronic relay (such as Model 7501), a 0-50 ° C electric contact mercury thermometer (model WXG) A matching air conditioner socket, installed in an electrical board, and then a good line. Because the air conditioner can be switched off after 2 minutes of shutdown, to make the electric contact mercury thermometer not too sensitive, a 2 cm long latex tube for blood transfusion should be put on the temperature sensing part.