我院于1995年购置MET-ASCA(美国)全自动生化分析仪,通过三年的应用,遇到一些故障,并摸索出了一些解决方法,现总结如下,供使用该仪器的同道参考。1 常见故障的原因与处理1.1 分析仪错误—温度失控1.1.1 检查开始和结束的TM值,如果开始值2000,则39号比色池的水温太低,如果TM
In 1995, our institute purchased the MET-ASCA (United States) automatic biochemical analyzer. After three years of application, it encountered some problems and developed some solutions. It is summarized as follows, for the reference of using the instrument. 1 Causes and Handling of Common Faults 1.1 Analyzer Errors - Temperature Runaway 1.1.1 Check the start and end TM values. If the start value is 2000, the water temperature of the Cuvette No. 39 is too low, if the TM