目的探讨FAME全自动酶免分析系统出现吐板现象的原因及相应对策。方法按设定程序在全自动加样系统加样后,将微板移至FAME全自动酶免分析系统,进行温育、洗板、显色、结果判读。结果8个月时间,FAME共出现9次吐板现象。结论严格遵守操作规程,做好标本抗凝、试剂槽内加入足量试剂并避免气泡产生、试验前将试剂槽条形码及微板板底擦干、对FAME进行维护保养等环节,可减少吐板现象的发生。
Objective To investigate the causes and corresponding countermeasures of the phenomenon of vomiting in FAME automatic enzyme immunoassay system. Method According to the setting program in the automatic loading system after loading, the microplate will be moved to FAME automatic enzyme analysis system for incubation, washing, color, the results of interpretation. Results 8 months, FAME appeared a total of 9 times vomit phenomenon. Conclusions Strict adherence to the operating procedures, good anticoagulation of specimens, addition of sufficient reagents in the reagent tank and avoidance of air bubbles. Drying of reagent tank barcodes and microplate bottoms before the test, maintenance of the FAME, etc. can reduce vomiting The occurrence of the phenomenon.