一个时期以来,北京医疗仪器维修中心不断地收到各地医院的来信,询问有关 AVL940血气分析仪在使用过程中所遇到的问题。维修人员到各个医院服务时,许多操作者也提出了各种疑难。除个别专门解答外,对一些带普遍性的问题,现向广大用户做一解答,同时也供其他读者参考。1.PO_2电极报警后经研磨处理一般能通过,但为什么 PCO_2电极研磨后性能更差?出现这个问题的用户多数是研磨 PCO_2电
A period of time, Beijing Medical Instrument Maintenance Center has received letters from hospitals around the country, asked about the AVL940 blood gas analyzer in the course of the problems encountered. When service personnel go to various hospital services, many operators also raise various problems. In addition to individual solutions, for some of the issues with the generality, is to make a answer to the majority of users, but also for other readers for reference. 1.PO_2 electrode after the alarm can generally be passed through the grinding, but why PCO_2 electrode grinding worse performance? The majority of users of this problem is grinding PCO_2 electricity