最近几年来,用硝酸钍容量法测定氟很被注意。为了确定滴定的终点,曾建议在缓冲溶液中应用各种不同的指示剂。采用直接电位滴定的试验未曾得到良好的结果,某些作者设计了复杂的电位滴定线图,并用低铁高铁电极(ферроферриэле-ктрод)作为指示电极。我们工作的目的是研究直接电位滴定氟离子的可能性,为此必须研究各种因素对于指示电极性能的影响。我们用光滑铂电极在普通电位滴定装置上进行试验。饱和甘汞半电池作为參比电极;盐桥用10%的硝酸铵溶液。在容器内注入50毫升含有1-15毫克氟离子的溶液,一面搅拌,一面
In recent years, the thorium nitrate volumetric method for the determination of fluorine is noticed. In order to determine the end of the titration, it has been suggested to use a variety of different indicators in the buffer solution. Experiments using direct potentiometric titration have not yielded good results. Some authors have designed complex potentiometric titration diagrams and used the low-iron high-iron electrode (ферроферриэле-ктрод) as the indicator electrode. The purpose of our work is to study the potential of direct potentiometric titration of fluoride ions, for which reason it is necessary to study the influence of various factors on the performance of the indicator electrode. We used a smooth platinum electrode in the ordinary potentiometric titration device for testing. Saturated calomel half-cell as a reference electrode; salt bridge with 10% ammonium nitrate solution. In a container, inject 50 ml of a solution containing 1 to 15 mg of fluoride ion while stirring it