Sysmex CS5100全自动凝血分析仪性能验证

来源 :血栓与止血学 | 被引量 : 0次 | 上传用户:skyboat521
下载到本地 , 更方便阅读
声明 : 本文档内容版权归属内容提供方 , 如果您对本文有版权争议 , 可与客服联系进行内容授权或下架
论文部分内容阅读
目的验证Sysmex CS5100全自动凝血仪的性能。方法从批内精密度及批间精密度、准确度、携带污染率、FDP线性范围共四个方面对Sysmex CS5100全自动凝血分析仪系统性能进行验证。结果精密度最大CV为6.55%;准确度最大偏倚为XXXX%;携带污染率最大值为-1.06%;FDP线性范围为0.954~5.967 g/L,相关系数为r为0.9992。结论 Sysmex CS5100全自动凝血分析仪具有较高精密度和准确度,携带污染率低,FDP线性范围宽,结果稳定、准确、可靠。 Objective To verify the performance of the Sysmex CS5100 fully automatic coagulometer. Methods The performance of the Sysmex CS5100 automatic coagulation analyzer system was validated in terms of intra-assay precision, inter-assay precision, accuracy, carryover contamination rate, and linear range of FDP. Results The maximum precision CV was 6.55%, the maximum accuracy was XXXX%, the maximum contamination rate was -1.06%, the linear range of FDP was 0.954 ~ 5.967 g / L, and the correlation coefficient was 0.9992. Conclusion Sysmex CS5100 automatic coagulation analyzer with high precision and accuracy, low carrier pollution rate, FDP linear range, the results of stable, accurate and reliable.