论文部分内容阅读
目的对乌鲁木齐市二宫街道社区卫生服务中心新建放射诊疗工作场所进行放射防护控制效果评价,确定放射防护设施和防护措施在控制职业照射和防止潜在照射方面的有效性、适宜性,保障放射工作人员和公众的健康与安全。方法将现场检测数据与国家标准进行比较。结果该中心工作场所周围区域空气比释动能率合格率为100%,影像质量检测结果为92.7%,放射诊疗工作场所放射防护控制效果基本达到国家标准规定要求,需在个人防护用品的配置上进一步完善,维修CT机,确保受检者及放射工作人员安全。结论该中心新建放射诊疗建设项目判定为职业病危害一般类的建设项目,该中心放射放射诊疗工作场所放射防护控制效果基本达到国家标准要求,能基本保障放射工作人员和公众的健康与安全。
Objective To evaluate the effectiveness of radiological protection and control of new radiation clinic in Nangong Community Health Service Center of Urumqi and to determine the effectiveness and suitability of radiological protection facilities and protective measures in controlling occupational exposure and potential exposure and to ensure that radiation workers And public health and safety. The method compares the field test data with national standards. Results The qualified ratio of air kerma rate in the surrounding area of the center was 100% and the image quality test result was 92.7%. The radiological control effect of radiotherapy and treatment work place basically met the requirements of national standards and required further configuration of personal protective equipment Improve and maintain the CT machine to ensure the safety of subjects and radiation workers. Conclusion The construction project of the new radiation clinic construction project of this center is judged as the general category of occupational hazards. The radiological control and prevention work in the radiology clinic basically meets the requirements of national standards and can basically guarantee the health and safety of radiation workers and the general public.