论文部分内容阅读
目前我军在野战条件下,多装备10—30mAX 线机。其特点是轻便、灵活。适合于野战作业。但因其容量小,平时极少使用,战时突然大工作量使用,常发生 X 线管损坏现象。为防止 X 线管损坏,长期停用的或新装的X 线机,使用前必须对 X 线管进行训练,以提高 X 线管真空度。我们在实践中发现,国产小型 X 线管,按传统训练方法,在训练过程中常出现 X 线管真空不良现象。或者训练结束后,试用时又出现X 线管真空不良现象。我们对小型 X 线管训练方法改进后,避免了上述现象。一、对1BEY1—75型 X 线管的试验F10型 X 线机配装国产1BEY1—75型X 线管,按传统方法训练(如表1所示)于65 kV时,发生 X 线管真空不良现象。
At present, our military in the field conditions, more equipment 10-30mAX line machine. It is characterized by light, flexible. Suitable for field operations. However, due to its small capacity, usually rarely used, the sudden heavy workload during the war, X-ray tube damage often occurs. To prevent damage to the X-ray tube, long-term disabled or newly installed X-ray machine, the X-ray tube must be trained before use to improve the X-tube vacuum. In practice, we found that domestic small X-ray tube, according to the traditional training methods, often appear in the training process X-ray tube vacuum bad phenomenon. Or after the training, trial X-ray tube vacuum appears bad phenomenon. We have avoided the above-mentioned phenomenon after the improvement of small X-ray tube training method. First, the test of 1BEY1-75 X-ray tube F10 X-ray machine equipped with domestic 1BEY1-75 X-ray tube, according to the traditional method of training (as shown in Table 1) at 65 kV, the occurrence of X-ray tube vacuum is not good phenomenon.