论文部分内容阅读
蔡司Theo 020光学经纬仪,采用弹片元件空气阻尼的自动整平装置,阻尼盒与阻尼塞之间间隙很小,但要求周围均匀,才不致相碰,因此调整颇感困难。用光隙法调整阻尼盒,可达到既简捷又理想的效果。现就此类结构用光隙法调整阻尼盒,作一简单介绍:1.用金属板做一块与020经纬仪左支架盖板同样大小的盖板作助调器(图1)。2.在φ9.2mm孔中装上一个6V、5W的灯泡,接上电线,注意灯泡顶至盖板的距离应小于14毫米,以免碰坏棱镜。
Zeiss Theo 020 optical theodolite, the use of shrapnel air damping automatic leveling device, damping the gap between the damper and the stopper is small, but requires around the uniform, it will not collide, so the adjustment quite difficult. Adjusting the damping box with the aperture method can achieve both simple and ideal results. Now on the structure of the aperture gap method to adjust the damping box for a brief introduction: 1. Use the metal plate to make an adjuster (Figure 1) with a cover of the same size as the 020 theodolite left bracket cover. 2. In φ9.2mm hole with a 6V, 5W bulb, connected to the wire, pay attention to the distance from the top of the lamp to the cover should be less than 14 mm, so as not to touch the prism.