国内中型诊断X线机中量限保护装置日趋晶体管化,其典型线路多数是沿用电子管阴极耦合双稳电路的形式改成晶体管开关线路中射耦双稳线路。该电路具有明显的幅度鉴别特性,工作点在截止区和饱和区上跳跃进行,所以一般讲是个较理想的一次性过载保护线路。但此电路本身存在的“回差”现象使摄片仟伏调节不便。一九七五年十月,全国200mA X线机质量检查测试会议规定,摄影仟伏过载数值应控制在±1KVP以内,超过此值机器将指示过载,自动切断摄影线路。当仟伏由高端调向低端时,由过载状态进入正常状态所允许
The domestic medium-sized diagnostic X-ray machine in the limit of protection devices increasingly transistor, the majority of its typical lines are used in the form of tube-cathode coupling bistable circuit into the transistor switching circuit in the radio-coupled bistable line. The circuit has obvious characteristics of amplitude discrimination, the work point in the cut-off area and the hopping on the saturation zone, so generally speaking is an ideal one-time overload protection circuit. However, this circuit itself, the “back to the poor” phenomenon so that the thousands of volts to adjust the inconvenience. In October 1975, the national 200mA X-ray machine quality inspection meeting stipulated that the photographic thousand-volt overload value should be controlled within ± 1KVP. If the value exceeds this value, the machine will indicate overload and cut off the photography line automatically. When thousands of volts from the high-end to the low-end, from the overload state to allow the normal state