我们对国产镍铬—镍硅和镍铬—考铜热电偶做过堆内辐照试验,现将结果扼要介绍如后。一、辐照前、后的电势值比较辐照前标定过的热电偶试样,在原子能所实验性重水反应堆9号中央垂直孔道内进行辐照,经热中子积分通量为1.6×10~(20)中子/cm~2,快中子(能量≥2.9兆电子伏),积分通量为4.5×10~(18)中子/cm~2的辐照考验后,将它们置于Y-300型校验分度用油炉内重新标定。结果表明,在室温至300℃校验温度范围内,辐照后热电偶的分度值,偏离标准分度表(JB-913-72)的最大值:镍铬—镍硅热电偶为3.2℃,镍铬—考铜热电偶为3.0℃,即都在标
We made domestic nickel-chromium-nickel-silicon and nickel-chromium-test copper thermocouple heap irradiation experiments done, the results are briefly described as follows. First, the potential before and after irradiation Comparison of pre-irradiation calibration of the thermocouple sample in the atomic energy of experimental heavy water reactor No. 9 central vertical aperture irradiation, the thermal neutron flux of 1.6 × 10 ~ (20) neutron / cm ~ 2, fast neutron (energy ≥ 2.9 MeV eV), and integrated flux 4.5 × 10 ~ (18) neutron / cm ~ Y-300-type calibration indexing furnace re-calibration. The results show that the division of the thermocouple after irradiation is deviated from the maximum value of the standard graduation table (JB-913-72) within the calibration temperature range from room temperature to 300 ℃: the Ni-Cr-Ni-Si thermocouple is 3.2 ℃ , Nickel-chromium - test copper thermocouple is 3.0 ℃, which are in the standard