目的比较新型静脉尿路造影压迫器与传统压迫器在普通剂量造影中的应用价值。资料与方法选择两组共76例患者,给予同等压力,同一压迫部位,分别使用新型压迫器与传统压迫器,进行静脉尿路造影,测量其在3 min、5 min、7 min和15 min加压时及30 min减压时,对比剂在肾盂及膀胱内的灰度值变化,并予以比较。结果采用新型压迫器的患者,在7 min和15 min时,肾盂内的对比剂灰度值高于采用传统压迫器的患者,而3 min、5min、7 min和15 min时,膀胱内对比剂灰度值比采用传统压迫器的患者低,提示新型压迫器的压迫效果优于传统压迫器。结论新型压迫器对输尿管的压迫效果,显著优于传统压迫器,有较高的临床应用价值。
Objective To compare the value of the new intravenous urography and traditional compressors in conventional dose contrast imaging. Materials and Methods A total of 76 patients in both groups were given the same pressure and the same compression site. The venous urography was performed using a new type of compression device and a conventional compression device. The changes were measured at 3 min, 5 min, 7 min and 15 min Pressure and 30 min decompression, the contrast agent in the renal pelvis and bladder gray value changes, and be compared. Results At the 7 min and 15 min, the gray value of the contrast agent in the renal pelvis was higher than that of the patients with the traditional compressors. At 3, 5, 7 and 15 minutes, the intravesical contrast agent Grayscale values were lower in patients with traditional compressors, suggesting that the new type of compressors performed better compression than traditional compressors. Conclusions The compression effect of the new type of compression device on the ureter is significantly superior to the traditional compression device, which has a high clinical value.