论文部分内容阅读
目的:调查血糖仪测定毛细血管血糖、静脉全血血糖与生化仪葡萄糖氧化酶法测定静脉血清糖之间的差别,了解血糖仪检测质量,规范血糖仪管理为临床治疗提供可靠数据。方法:血糖仪检测末梢血、静脉血;用全自动生化分析仪葡萄糖氧化酶法测试静脉血清血糖值进行比对试验,并对其结果进行分析。结果:血糖仪测定末梢血、静脉血、生化仪测定静脉血清的结果分别为(4.64±1.87)mmol/L、(4.69±1.61)mmol/L、(3.87±0.92)mmol/L,统计学检测三组结果无显著意义(P>0.05)。结论:血糖仪检测质量总体上是可靠的,医疗护理中可根据病情选择适当的血糖监测方法。
Objective: To investigate the difference between capillary blood glucose, venous whole blood glucose and glucose oxidase method to determine venous serum sugar, to understand the quality of blood glucose meter, and to regulate the management of blood glucose meter to provide reliable data for clinical treatment. Methods: Peripheral blood and venous blood were measured by blood glucose meter. The blood glucose level was measured by glucose oxidase method with automatic biochemical analyzer, and the results were analyzed. Results: The blood glucose and venous blood were measured by blood glucose meter and the results of biochemical analyzer were (4.64 ± 1.87) mmol / L, (4.69 ± 1.61) mmol / L and (3.87 ± 0.92) mmol / L respectively No significant difference was found between the three groups (P> 0.05). Conclusion: The quality of blood glucose meter test is generally reliable, medical care, according to the disease to choose the appropriate blood glucose monitoring methods.