目的明确Wright雾化器在雾化过程中的释雾量变化,为建立支气管激发试验质量控制标准提供依据。方法对4个不同的Wright雾化器分别加入2mL生理盐水,在压缩空气(压力=344.7kpa)的驱动下,以8、9、10L/min的流速分别雾化2分钟各5次,分别测定其释雾量并进行组间比较,并计算其变异率。结果 4个不同Wright雾化器以8L/min的流量雾化2分钟的平均释雾量分别为:0.1mL/min、0.096mL/min、0.095mL/min、0.108mL/minm(p<0.05),变异率分别为:0%、10%、0%、4.54%;以9L/min的释雾量分别为:0.125mL/min、0.124mL/min、0.098mL/min、0.130mL/min(p<0.05),变异率分别为:7.69%、7.69%、5%、7.4%;以10L/min的释雾量分别为:0.142mL/min、0.141mL/min、0.122mL/min、0.137mL/min(p<0.05),变异率分别为:10%、6.89%、7.69%、7.14%。结论不同Wright雾化器在同一压缩空气流速下的释雾量的差异有显著性,同一Wright雾化器在不同压缩空气流速下的释雾量有一定的变化,但变异率不超过10%。因此采用Wright雾化器进行2分钟潮气呼吸法支气管激发试验前,应对不同的雾化器分别进行释雾量定标,以保证达到要求的释雾量,相同的雾化器在同一流速下产生的释雾量相对稳定。
Objective To clarify the change of defogging fog of Wright nebulizer in the process of atomization to provide the basis for establishing the quality control standard of bronchial provocation test. Methods Four different Wright nebulizers were respectively added with 2 mL of normal saline, and were respectively aerosolized for 2 minutes at the flow rates of 8, 9 and 10 L / min under the action of compressed air (pressure = 344.7 kPa) The release of fog and compare between groups, and calculate the mutation rate. Results The average amount of defogged fogging of 4 different Wright nebulizers at a flow rate of 8 L / min for 2 minutes was 0.1 mL / min, 0.096 mL / min, 0.095 mL / min, 0.108 mL / minm, , The rate of variation was 0%, 10%, 0% and 4.54%, respectively. The amount of released fog at 9 L / min were 0.125 mL / min, 0.124 mL / min, 0.098 mL / min and 0.130 mL / <0.05), the rate of variation was 7.69%, 7.69%, 5% and 7.4%, respectively. The amount of released fog at 10 L / min was 0.142 mL / min, 0.141 mL / min, 0.122 mL / min and 0.137 mL / min (p <0.05). The mutation rates were 10%, 6.89%, 7.69% and 7.14%, respectively. Conclusions There is a significant difference in the amount of defogging mist between different Wright nebulizers under the same compressed air flow rate. The same Wright nebulizer has a certain amount of defogging mist at different compressed air velocities, but the variation rate does not exceed 10%. Therefore, before using the Wright nebulizer for 2 minutes of tidal breathing bronchial provocation test, different nebulizers should be calibrated respectively to ensure that the required amount of defrosting is achieved. The same nebulizer is generated at the same flow rate The release of fog is relatively stable.