目的:探讨螺旋CT扫描时降低管电流对儿童寰枢椎病变图像质量的影响。方法:34例疑有寰枢椎病变的病例随机分成低剂量和常规剂量组,分别用80mA、180mA管电流行螺旋CT扫描。记录容积CT剂量指数(CTDIvol)、扫描长度,计算出平均剂量长度乘积(DLP)。对不同扫描参数的图像质量分别进行评分,并进行统计学分析。结果:低剂量(80mA)和常规剂量(180mA)的CTDIvol分别为8.7mGy和19.6mGy,DLP分别为(36.80±3.60)mGycm和(82.14±7.18)mGycm,各组可满足临床诊断要求的图像比例无统计学差异。结论:儿童寰枢椎螺旋CT检查时降低管电流,既明显减少受检者的辐射剂量,又能保证图像质量,满足临床诊断的要求。
Objective: To investigate the effect of decreasing the tube current on the image quality of atlantoaxial lesion in children with spiral CT. Methods: Totally 34 cases of suspected atlantoaxial diseases were randomly divided into low dose group and conventional dose group. They were performed spiral CT scan with 80mA and 180mA respectively. The volume CT dose index (CTDIvol) was recorded and the length of the scan recorded to calculate the average dose length product (DLP). The image quality of different scanning parameters were scored separately and statistically analyzed. Results: The CTDIvol of low dose (80mA) and routine dose (180mA) were 8.7mGy and 19.6mGy, and the DLP were (36.80 ± 3.60) mGycm and (82.14 ± 7.18) mGycm, respectively, No statistical difference. CONCLUSION: The reduction of the tube current during pediatric atlantoaxial spiral CT examination can not only significantly reduce the radiation dose but also ensure the image quality and meet the requirements of clinical diagnosis.