【摘 要】
:
1故障现象开15MVX线时,DORATE2和DOR-TE1不同步,DORATE2异常增高,操作台显示“3.MONITORSYNC(SW)”和“22.HIRATE2”连锁。6MVX线和电子线出束正常。2故障分析根据DORATE1和D
【机 构】
:
山东省肿瘤医院,山东省肿瘤医院,山东省肿瘤医院,山东省肿瘤医院 济南市250117,济南市250117,济南市250117,济南市250117
论文部分内容阅读
1故障现象开15MVX线时,DORATE2和DOR-TE1不同步,DORATE2异常增高,操作台显示“3.MONITORSYNC(SW)”和“22.HIRATE2”连锁。6MVX线和电子线出束正常。2故障分析根据DORATE1和DORATE2信号流程图(附图)可知,6MVX线和电子线出束正常,说明X线电离室和E线电离室正常,PC2板和PC4板也正
1 Symptom When the 15MVX line is on, DORATE2 and DOR-TE1 are not synchronized and DORATE2 is abnormally high. The console displays “3.MONITORSYNC (SW)” and “22.HIRATE2”. 6MVX wire and electron beam normal beam. 2 Fault Analysis According to the DORATE1 and DORATE2 signal flow chart (with photos) shows, 6MVX line and electron beam out of the normal, indicating that the X-ray ionization chamber and the E-line ionization chamber is normal, PC2 board and PC4 board is also positive